AMI氧氣傳感器
AMI Model Number | T-2 | T-4 |
Max Range | 0-10,000ppm * | 0-10,000ppm * |
Min Range | 0-5ppm | 0-5ppm |
Min Detection Limit | 0.05ppm | 0.05ppm |
90% full scale response time (increasing oxygen levels) | 8 sec | 8 sec. |
1 minute air exposure to <10ppm come-down time | <15 minutes at 70°F | <15 minutes at 70°F |
Life Expectancy | 12-24 months In <100ppm oxygen | 12-24 months In <100ppm oxygen |
Warranty F.O.B. Costa Mesa, CA | 6 months | 6 months |
Application | PPM oxygen with up to 10ppm H2S. Virtually all background gases including carbon dioxide from zero up to 99%. | PPM oxygen with up to 500ppm H2S. Virtually all background gases including carbon dioxide from zero up to 99%. |
AMI Model Number | P-2 | P-3 | P-4 | P-5 |
Max Range | 0-50% | 0-25% | 0-99% | 0-25% |
Min Range | 0-1000ppm | 0-1000ppm | 0-1% | 0-1000ppm |
Min Detection Limit | 10ppm | 100ppm | 1000ppm | 100ppm |
90% full scale response time (increasing oxygen levels) | 8 sec | 30 sec. | 30 sec | 30 sec |
1 minute air exposure to <10ppm come-down time | N/A | N/A | N/A | N/A |
Life Expectancy | 12-15 months In 20.9% 0xygen | 9-12 months In 20.9% 0xygen | 18 months In 20.9% 0xygen | 9-12 months In 20.9% 0xygen |
Warranty F.O.B. Costa Mesa, CA | 6 months ** | 6 months ** | 6 months ** | 6 months ** |
Application | Percent oxygen in inert or reducing background gas | Percent oxygen in CO2 containing background gas | Percent oxygen in inert background gas | Percent oxygen in CO2 containing background gas with up to 500ppm H2S |