Specifications |
| 3350 | 5350 |
Size Range: | 0.3 - 25.0μm | 0.5 - 25.0μm |
Standard Channel Sizes: | 0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm | 0.5, 0.7, 1.0, 3.0, 5.0, 10.0μm |
Optional Channel Sizes: | 0.3, 0.5, 1.0, 5.0, 10.0, 25.0μm | 0.5, 1.0, 3.0, 5.0, 10.0, 25.0μm |
| | |
Flow Rate: | 100.0 LPM (3.53 CFM) |
Counting Efficiency: | 3350, 50% @ 0.3μm; @ sizes > 0.45μm; 5350, 50% @ 0.5μm; @ sizes > 0.75μm Per JIS B 9921 and ISO 21501-4 |
Light Source: | Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF) |
Zero Count Level: | <1 count / 5 minutes (per JIS B 9921) |
Concentration Limits: | 400,000 Particles/ft3 (14,120,000/M3) @10% Coincidence Loss, Suitable for ISO classes 1 through 8 Cleanrooms |
Calibration: | Meets ISO 21501-4 Calibration Standard |
Count Modes: | Automatic, Manual, Beep, Concentration, Cumulative/Differential and Real Time (with LMSNet or LMS Express RT) |
Data Storage: | 3,000 Sample Records (Rotating Buffer), which includes particle and environmental data with locations and times |
Communication Modes: | USB; Ethernet; RS-485; Secure Data Transfer to and Storage on USB Flash Drive |
Supporting Software: | LMS XChange, LMS Express and Express RT; LMS Standard; LMS Pharma; CREWin |
Environmental Sensors: | Optional (up to 4): Temperature/Relative Humidity, Air Velocity, Differential Pressure or others |
Touch Screen Display: | 5.7-inch (14.47 cm) color, 320x240 |
Location Labels: | Up to 400, based on Location Label Length |
Printer: | Built-in Thermal Printer (non-printer model available) |
Reports: | Pass/Fail Report Wizard for ISO-14644-1, EU GMP Annex 1 and FS-209 Report Formats |
Alarm: | Integrated Alarm Buzzer and External Alarm Connector. Alarms on Counts, Low Battery, Sensor Failure |
Enclosure: | Stainless Steel |
Sample Output: | Internally filtered to HEPA standards (>99.997% @ 0.3μm) |
Vacuum Source: | Internal Closed-loop Controlled Pump: allows the use of 3/8” ID Sample Tubing of up to 12 Meters (40 feet) @ 100 LPM |
Power: | 100-250 VAC, 47-63 Hz |
Battery: | Li-Ion, Removable & Rechargeable; Typical Battery Operation ~4 hours*; Tyical Recharge Time ~4 hours |
Dimensions: | 14.65” (l) x 8” (w) x 10.4” (h) [37 x 20 x 26 cm] |
Weight: | With Battery: 26 lbs (11.8 kg) Without Battery: 23 lbs (10.43 kg) |
Languages: | English, Fran?ais, Deutsch, Italiano, Espa?ol |
Environmental Conditions |
Operating: | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
Storage: | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
Accessories |
Included: | Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Power Cable and Spare Fuse; LMS XChange Software; Printer; Printer Paper |
Optional: | Wireless Network Adapter; Removable Rechargeable Li-Ion Battery; Carrying Case with Wheels; Temperature/Relative Humidity Probe; Differential Pressure Probe; Intelligent Filter Scanning Probe; Air Velocity Probe; Spare Removable Rechargeable Li-Ion Battery; External Battery Charger; LMS Express, RT Software; LMS Pharma Edition Software; LMS Standard Software; Validation Documentation; Printed Operating Manual; Cleanroom Cart |